• Unit: The measurement unit is displayed during measurement.
• All CH:
Sets the Input, Range, and Filter values to the same values for all the channels in the
displayed group. Click the "Set" button next to "All CH" to make the settings.
(2) Logic/Pulse Settings
Device Number: Specify which of the connected devices to use for logic input.
Logic: Off, On
Select On if a Device Number was specified.
• Line Color: Specifies the colors for channels 1 to 4.
Pulse
• Line Color:
Specifies the color.
• Input:
OFF, Counts, Revolutions, Inst.
OFF: Pulses are not measured.
• Revolutions: Counts the number of pulses at 1-second intervals and then multiplies them by
a factor of 60 to display them as revolutions.
• Counts:
Displays the cumulative number of pulses for each sampling interval from the
start of measurement.
• Inst.:
Counts the number of pulses for each sampling interval.
Range: Can be set for each of the modes.
Mode
Revolutions
Counts
Inst.
Slope: Sets the condition for the input signal operation.
• H: Operates when the signal is a rising signal
• L: Operates when the signal is a falling signal.
Range
50k, 500k, 5M, 50M, 500M C/F.S.
500, 5k, 50k, 500k RPM/F.S.
50k, 500k, 5M, 50M C/F.S.
Maximum number of pulse inputs
50 k/sampling interval
50k/s
50 k/sampling interval