Software
4.8 Measurement Parameters Settings
Click the "Settings" button on the main menu to enable device and measurement parameters for the GL450
to be made.
CHECKPOINT
The GL450 settings are changed whenever the "OK" or "Apply" button is clicked in each setting menu.
AMP Settings
(1) Select CH
Specify a group or groups of channels in the range [1 to 10] through [91 to 100].
Analog AMP Settings:
• Line Color: Specifies the color for each channel.
• Annotation: Up to 11 characters can be input for each channel.
• Input: OFF, DC, Temp, RH
• Range
• Filter: Off, On
4-16
CAUTION
Certain characters used here may not be displayable on the GL450 monitor.
Sets the input type for each channel.
CHECKPOINT
For RH, the following settings are automatically made:
DC: 1V; Scaling: 0V → 0%, 1V → 100%
DC: 20, 50, 100, 200, 500 mV, 1, 2, 5, 10, 20, 50 V
CHECKPOINT
For the 1-5V range, select 5V, and then specify 5V for the upper span limit and 1V for the
lower span limit.
Temp: Specify one of the following:
TC-K, TC-J, TC-T, TC-E, TC-B, TC-R, TC-S, TC-N, TC-W, Pt100, JPt100.
Select On if you need to cut noise.