GRAPHTEC GL450 User Manual page 52

Midi logger
Hide thumbs Also See for GL450:
Table of Contents

Advertisement

r r r r r ALM Settings Window
The Alarm and Trigger settings are made here.
Setting
Combination
OR, AND
Repeat Trigger
On, Off
Mode
H, L, Win In, Win Out
Lower - Level - Upper
Numeric settings
Output
1 to 4
Pulse
• Mode
L, H, Win In, Win Out
• Lower - Level - Upper Numeric settings
• Output
1 to 4
Logic
• Mode
X, L, H
• Output
1 to 4
Trigger
• Start
Off, Level, Alarm, Ext., Specified Time
Alarm output selection: 1 to 4
• Stop
Off, Level, Alarm, Ext., Time, Specified Time
Alarm output selection: 1 to 4
Time setting: 1 s to 9999 h
Combination
Determines the alarm output conditions. The logic alarm cable (B-513, option) is required for alarm
output.
OR: Outputs an alarm when each alarm condition is met.
AND: Outputs an alarm when all of the alarm conditions are met.
Repeat Trigger
Used to specify repeated measurement. Measurement is performed repeatedly according to the
Trigger Start and Stop conditions until the START/STOP key is pressed.
On: Repeated measurement
Off: Single measurement operation
CHECKPOINT
If repeated measurement has been specified and Memory selected as the capture destination, data
captured to the GL450's internal memory is overwritten.
Selections available
Settings and Measurement
Setting method
ENTER→Select→ENTER
ENTER→Select→ENTER
ENTER→Select→ENTER
ENTER→Select→ENTER
ENTER→Select→ENTER
ENTER→Select→ENTER
ENTER→Select→ENTER
ENTER→Select→ENTER
ENTER→Select→ENTER
ENTER→Select→ENTER
ENTER→Select→ENTER
ENTER→Select→ENTER
ENTER→Select→ENTER
ENTER→Select→ENTER
ENTER→Select→ENTER
3-17