Settings and Measurement
Statistical Calculation .......... Statistical calculation is performed on the replay data between the
3-30
Alarm Slope: Specify the generation condition for the alarm search.
• Both: Both rising and falling
• Hi:
Rising only
• Lo:
Falling only
Next Alarm Match/Prev. Alarm Match:
Depending on whether Next or Prev. has been selected, matching
data from the current cursor position to the next alarm position, or
from the previous alarm position to the current cursor position is
searched for and displayed.
cursors.
Function: There are five types of between-cursor calculation
functions, and four of these can be selected.
• Off:
Calculation is not performed.
• Average: The simple average value of the data between the cursors
is displayed.
• Max:
The maximum value of the data between the cursors is
displayed.
• Min:
The minimum value of the data between the cursors is
displayed.
• Peak: The peak value of the data between the cursors is displayed.
• RMS: The effective value of the data between the cursors is
displayed.
2
R.M.S =
ΣD
/n
D: data N: number of data
Channel specification:
Specifies the two channels "A" and "B" for calculation. Press
EXECUTE to display the between-cursor calculation results.