GRAPHTEC GL450 User Manual page 60

Midi logger
Hide thumbs Also See for GL450:
Table of Contents

Advertisement

Alarm Cancel ...................... When On has been selected for the Alarm Hold parameter, select
Start/Stop Confirmation Message
.................................. If On has been selected for this parameter, a confirmation message
Power On Start ................... Initiates measurement as soon as the GL450 is turned on.
Room Temp. Compensation
.................................. This parameter enables room temperature compensation settings
AC Line Frequency ............. Select the setting that matches the frequency of the AC line used. Use
Burn Out ............................. This parameter enables or disables the function that moves to full
Temp. Unit ........................... This parameter specifies the temperature unit.
Alarm Cancel to cancel the alarm status when data capture has
stopped. To cancel the alarm status during a data capture operation,
press the MENU key and then the ENTER key.
box is displayed at the start of data capture. Select On to prevent any
misoperation during data capture.
Disable: Disables the Power On Start function.
Enable: Enables the Power On Start function.
when thermocouples are used. You can select from either Internal or
External room temperature compensation.
Internal: The GL450's room temperature compensation settings are
used.
External: Select this parameter when measuring compensation other
than that of the GL450.
Note: The Internal setting is normally used.
this parameter if there is AC line noise present when performing
measurement.
50 Hz, 60 Hz
CHECKPOINT
With the 50 Hz and 60 Hz frequencies, the sampling speed at which
AC line noise can be eliminated depends on the number of channels
installed (please see the following table for details).
Sampling Speed/Number of Channels Correlation Table (When the
Digital Filter is Enabled)
50/60: Can be eliminated at 50/60 Hz
100: Can be eliminated at 100 Hz or higher
NG:
Cannot be eliminated
–:
Cannot be selected
100 ms
10 CH
NG
20 CH
30 CH
40 CH
50 CH
60 CH
70 CH
100 CH
scale to inform of a sensor burnout in a thermocouple.
Off: The burnout function is disabled.
On: The burnout function is enabled.
Note: We recommend that On be normally used.
°C: Celsius
°F: Fahrenheit
If °F is selected, the scaling function is compulsorily enabled.
Sampling speed
200 ms
500 ms
1 s
100
50/60
50/60
NG
100
50/60
NG
100
NG
100
100
NG
NG
NG
Settings and Measurement
2 s
5 s
50/60
50/60
50/60
50/60
50/60
50/60
50/60
50/60
50/60
50/60
100
50/60
100
50/60
100
50/60
3-25