"FlexPrint" Printout Function
Purpose
The YAD02IS "Nice Label Express"
software from Sartorius lets you load
user-defined label printing formats and
the corresponding print instruction files
in your scale. This software lets you
connect any of a number of printers,
equipped with a variety of printer fonts,
to the RS-232 interface on your scale.
Features
With the "FlexPrint" option activated:
– Print command generates configured
printout (if print instruction file exists;
see table, next page). Print command
generates default printout (if print
instruction file does not exist).
– The function that generates an
automatic printout upon initialization
of an application cannot be used.
Initialization data can be output
only to a print instruction file.
The following items are output only
as standard printouts:
– Calibration/adjustment
– SETUP printouts
– p key in "Identifier:" display page
– p key on "Product data memory:"
display page
Print events
Printouts generated using the "Nice Label Express" software are divided into the following groups:
Event
1. p key with individual values
2. p key with text input
3. GLP /GMP header
4. GLP /GMP footer
5. Results, Application 1
6. Results, Application 2
7. Results, Application 3
8. Components, Application 1
9. Components, Application 3
– p on "Parameter" display page
for "Checkweighing" application:
print function carried out
– MR function carried out when p is
pressed during evaluation (info window)
in Totalizing and Statistics applications.
– To recall the file names, software ID
and version numbers, for FlexPrint, see
the section on "Basic Settings," Info
Display."
– Printout for legal metrology:
Weight blocks (special weight value
formats that are acceptable in legal
metrology) are designated by the
following line, which is printed in
both header and footer and cannot
be edited:
"------ CE [M] ------"
Examples of Weight Block Printouts
Without tare:
------ CE [M] -------
N
+
348.65 kg
------ CE [M] -------
Explanation
Print key
Input and p key
GLP header
GLP footer
Animal weighing, MR-CF
OK values, time-controlled print
MR, MR-CF
M+ printout
M+/M– printout
With tare:
------ CE [M] -------
G
+
459.70 kg
N
+
348.65 kg
T1 +
111.05 kg
------ CE [M] -------
Scale tare (2nd tare memory):
------ CE [M] ------
G
+
124.45 kg
N
+
100.00 kg
T1 +
24.00 kg
T2 +
0.45 kg
------ CE [M] ------
Preparation
§ Turn on the scale: press the e key
§ Configure FlexPrint in Setup:
press the H key
Printout
§ Select
: press
3 times, and then
soft key
O
§ Confirm
Application-
defined output
:
press
soft key
O
§ Select
FlexPrint
: press
key 3 times and then
soft key
O
§ Select
On
: press
soft key and
q
then
soft key
O
§ Save settings and exit Setup:
press
soft key
oo
File name for event group:
PPRINT
PDIRECT
PGMPHEAD
PGMPFOOT
PA1RES
PA2RES
PA3RES
PA1COMP
PA3COMP
soft key
q
soft
q
93