Sartorius FC Operating Instructions Manual page 53

Factory series electronic precision scales for hazardous areas/locations
Table of Contents

Advertisement

Examples
Practical Example: Weighing in Percent with Reference Weight Taken from Weight on Scale
Settings (changes in the factory settings required for this example):
Setup: Application parameters: Application 1: Weighing in percent
Setup: Printout: Application-defined output: Autoprint upon initialization: All values
Step
1. Delete previous setting if necessary
2. Prepare a container
for the parts
3. Tare the scale
4. Place the reference weight
on the scale
(here: 1821.48 g = 100%)
5. Initialize the scale
6. Unload the scale
7. Determine the percentage
of an unknown weight
8. If desired, print percentage
(here: 98.37%)
Key (or instruction)
c
Place the empty container
on the scale
t
Place weight equal
to reference weight in
the container
Start
soft key
Remove reference
sample from the container
Place sample to be
measured in the container
p
Display/Output
pRef
+
Wxx%
+
1821.48 g
P
98 37 %
Prc
+
98.37 %
100 %
53