Sartorius FC Operating Instructions Manual page 74

Factory series electronic precision scales for hazardous areas/locations
Table of Contents

Advertisement

Totalizing
L
Practical Example
Totalize counted pieces
Settings (changes in the factory settings required for this example):
Setup: Application parameters: Application 1: Counting
Setup: Application parameters: Application 2: Off
Setup: Application parameters: Application 3: Totalizing: Evaluated values: Net + calculated
Setup: Application parameters: Application 3: Totalizing: Evaluation mode, MR function: Final evaluation, display + print
Step
1. Turn on the scale and configure
the settings as indicated above
2. Delete old totalization data, if necessary
3. Tare the scale
4. Toggle to Application 1: Counting
5. Place the displayed number of
parts on the scale (here: 10 pcs)
6. Initialize the Counting application
7. Remove the reference sample
quantity and toggle to Totalizing
8. Place a number of parts on the
scale (here: 50 pcs)
74
Press key(s)
(or follow instructions)
e
c
t
D
Place parts to be counted
on the scale
Start
soft key
Unload the scale
D
Place parts on the scale
Display/Output
6
isoTST
6
isoTST
6
isoTST
nRef
wRef
35.28900 g
6
isoTST
6
isoTST
10 pcs