Sartorius FC Operating Instructions Manual page 71

Factory series electronic precision scales for hazardous areas/locations
Table of Contents

Advertisement

Practical Example
Document the evaporated amount of a sample with defined surface, temperature and air pressure at preset intervals of 1 minute, 30 seconds.
Settings (changes in the factory settings required for this example):
Setup: Application parameters: Application 2: Time-controlled functions
Setup: Balance/scale functions: Taring: Without stability
Setup: Printout: Application-defined output: Stability parameter: Without stability
Step
1. Turn on the scale and configure
the settings as indicated above
2. Delete stored values, if necessary
3. Place container with sample
on the scale and tare
4. Enter time interval: 1 minute,
30 seconds
5. Store time interval
6. Begin documentation
(Time remaining until the
next printout is displayed
in the text line)
Printout of evaporated amount
every 1 minutes
7. Stop the documentation procedure
Press key(s)
(or follow instructions)
e
c
t
1 . 3 0
Interv
soft key
Start
soft key
Stop
soft key
Display/Output
71