ISO/GMP-compliant
Printout/Record
You can have the parameters pertaining
to weighing conditions printed before
(GMP header) and after (GMP footer)
the values from the weighing series.
These parameters include:
– Date
– Time at the beginning of a weighing
series
– Scale manufacturer
– Scale model
– Model serial number
– Software version
– Lot number (weighing series no.)
– Time at the conclusion of the weighing
series
– Field for operator signature
Operating the Scale with an ISO/GMP-
capable Documentation Device (Printer)
ISO/GMP-compliant documentation
requires a computer with special
software. Contact Sartorius for a
detailed description for creating this
software.
Setting:
Setup: Printout: ISO/GMP printout:
Always on
The record is output to a Sartorius
YDP03-0CE Data Printer or a computer.
End GMP printout:
Press c
End GMP printout while application
is active:
This requires the following settings:
Setup: Device: Keys: CF function in
application: Clear only selected
applications
Press c
> Text line: CF selected: clear application
GLP
Press the
soft key
100
--------------------
17.01.1997
16:12
SARTORIUS
Mod.
FC6CCE-HX
Ser. no.
60419914
Ver. no.
01-35-18
ID
12345678901234
--------------------
L ID
12345678901234
nRef
10 pcs
wRef
1.35274 g
Qnt
+
235 pcs
Qnt
+
4721 pcs
S ID
12345678901234
Qnt
+
567 pcs
--------------------
17.01.1997
16:13
Name:
--------------------
--------------------
17.01.1997
16:24
SARTORIUS
Mod.
FC6CCE-HX
Ser. no.
60419914
Ver. no.
01-35-18
ID
--------------------
L ID
12345678901234
Internal calibration
Start:
manual
Diff. +
0.006 g
Internal calibration
completed
Diff. +
0.000 g
--------------------
17.01.1997
16:25
Name:
--------------------
Dotted line
Date/time
Scale manufacturer
Scale model
Scale serial number
Software vers. (display and control unit)
Scale ID no.
Dotted line
Weighing series no.
Application initialization value
Application initialization value
Counting result
Counting result
ID for counting result
Counting result
Dotted line
Date/time
Field for operator signature
Blank line
Dotted line
Record of Internal
Calibration/Adjustment:
Dotted line
Date/time
Scale manufacturer
Scale model
Scale serial number
Software vers. (display and control unit)
Scale ID no.
Dotted line
Weighing series no.
Calibration adjustment mode
Beginning mode for calibration/adjustment
Difference after calibration/adjustment
Confirmation of completed
calibration/adjustment routine
Difference between current and target
values after calibration
Dotted line
Date/time
Field for operator signature
Blank line
Dotted line