Sartorius FC Operating Instructions Manual page 68

Factory series electronic precision scales for hazardous areas/locations
Table of Contents

Advertisement

Checkweighing
H
Practical Example
Checkweighing samples of 170 g, with an allowable tolerance of –5 g and +10 g. Printout of upper and lower tolerance limits.
Weighed values are printed out automatically when stability is reached and the weight value is within the checkweighing range.
Settings (changes in the factory settings required for this example):
Setup: Application parameters: Application 2: Checkweighing: Automatic printout of OK values: On
Step
1. Turn on the scale and configure
the settings as indicated above
2. Delete previous values, if necessary
3. Prepare a container
for the samples
4. Tare the scale
5. Enter initialization values
6. Accept target value
(in this example: 170 g)
7. Save target and unload
the scale
8. Enter value for lower limit
(170 g – 5 g) and save
68
Press key(s)
(or follow instructions)
e
c
Place empty container
on the scale
t
Param.
soft key
Place ideal sample
in container
soft key
l
Remove ideal sample
from the scale
1 6 5
soft key
l
Display/Output