Extra Functions (In The Application Menu); Second Tare Memory (Preset Tare); In The Application Menu Second Tare Memory (Preset Tare - Sartorius FC Operating Instructions Manual

Factory series electronic precision scales for hazardous areas/locations
Table of Contents

Advertisement

Extra Functions (in the Application Menu)

Second Tare Memory
(Preset Tare)
Purpose
With this function, you can store the
weight currently on the scale as a tare
weight, or use the numeric keys to enter
a number for a preset tare weight.
You can use this function in
combination with a program from
Application 1 (such as counting,
weighing in percent), one from
Application 2 (checkweighing, time-
controlled functions) and one from
Application 3 (totalizing, formulation,
statistics) as well as with the extra
functions.
Features
– Store a weight on the scale in the
second tare memory (without numeric
input)
– Store a numeric value in the second
tare memory (input using the numeric
keys)
– Identify a net value as s when there
is a value stored in the second tare
memory
– You can assign this function to the
fourth or fifth soft key (from the right);
i.e., F4 or F5.
The soft key designation for this
PT1/T1
function is:
– Optional configuration in the Setup
menu for storing the current weight
readout as the container tare weight.
Any load subsequently placed on the
scale that is more than 70% of the tare
weight is automatically recognized
as a container and the scale is tared
automatically.
– Automatic printout when a value is
stored or input (see "Configuration")
– Press c to delete the (preset) tare
value
84
Factory Settings
No
Container tare weight:
Off
Automatic printout:
Soft Key Functions
PT1/T1
Store weight as tare value
PT1
Store input value
Preparation
Turn on the scale: press e
> The Sartorius logo is displayed; a self-test is performed
Select Extra function (F4) or Extra function (F5) in the Setup menu: press H
Application parameters
Select
then the
O
soft key once
Extra function (F4)
Select
press the
soft key 3+ (or 4+), then press the
q
2nd tare memory
Select
2nd tare memory
Confirm
2nd
Container tare
tare memory
weight
Automatic printout
ο = factory setting
see also the "Application Parameters (Overview)" in the chapter entitled "Configuration"
Save settings and exit the Setup menu: press the
Second Tare Memory in Legal Metrology
(i)PT1
– Press the
soft key to enter information about the tare value using the
number keys.
– The PT1 tare value is printed out with the net value.
Printout of the Data in the
2nd Tare Memory
The printout shows either:
– Net value N1,
– Tare weight T1, or
– Manually entered tare value PT1
N1
63.48 g
T1
138.73 g
PT1
150.00 g
N1:
Net weight (value) when a weight
is stored in the tare memory
T1:
Tare weight
PT1: Preset tare value entered using
the numeric keys
: press the
soft key 2+,
q
Extra function (F5)
or
soft key once
O
ο No
Yes
Net value
Tare/preset
ο Off
soft key
oo
:

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents