Animal Weighing
Practical Example
Determining Animal Weight With Automatic Start of 20 Subweighing Operations for Averaging; Automatic Printout of the Number of Sub-
weighing Operations and of the Animal Weight
Settings (changes in the factory settings required for this example):
Setup: Application parameters: Application 1: Animal weighing: Animal activity: Active
Setup: Application parameters: Application 1: Animal weighing: Decimal places in calculated result display: 2 decimal places
Setup: Application parameters: Application 1: Animal weighing: Printout: Average and calculated values
Setup: Printout: Application-defined output: Auto print upon initialization: All values
Step
1. Delete previous setting if necessary
2. Prepare a container (cage)
3. Tare the scale
4. Enter number of subweighing
operations for averaging
5. Save number
6. Weigh the first animal
7. Start automatic animal weighing
The scale delays starting the
subweighing operation until
three successive subweights lie
within the range defined for
an "active" animal
56
C
Key (or instruction)
c
Place empty cage
on the scale
t
2 0
mDef
soft key
Place 1st animal in cage
Start
soft key
When this criterion is met, the
subweighing series begins
Display/Output
weight value fluctuates due to animal activity