Example 3: Capture The Single Signal - Owon SDS6062 User Manual

Sds series, smart digital storage
Hide thumbs Also See for SDS6062:
Table of Contents

Advertisement

6. Demonstration
(8) Read the peak-to-peak voltages of Channel 1 and Channel 2 from the bottom left
of the screen (See Fig.6-2).
(9) Calculate the amplifier gain with the following formulas.
Gain = Output Signal / Input signal
Gain (db) = 20×log (gain)

Example 3: Capture the Single Signal

The digital storage oscilloscope takes the lead in providing the convenience capturing
of such non-periodic signals as pulse and burr, etc. If you intent to capture a single
signal, you can not set the trigger level and the trigger edge unless you have particular
priori knowledge of this signal. For example, if the pulse is the logic signal of a TTL
level, the trigger level should be set to 2 volts and the trigger edge be set as the rising
edge trigger. If it is uncertain as to the signal, you can make an observation of it in
advance under the automatic or ordinary mode to determine the trigger level and the
trigger edge.
The operation steps are as follows:
(1) Set the probe menu attenuation coefficient to 10X and that of the switch in the
probe to 10X (see "How to Set the Probe Attenuation Coefficient" on P14).
(2) Adjust the VOLTS/DIV and SEC/DIV knobs to set up a proper vertical and
horizontal ranges for the signal to be observed.
(3) Press the Acquire button to display the Acquire menu.
(4) Press the H1 button to display the Acquire Mode menu.
73
Fig.6-2 Waveform of Gain Measurement