State:ac - HP E1340A User Manual

Arbitrary function generator
Hide thumbs Also See for E1340A:
Table of Contents

Advertisement

:STATe:AC

CAUTION
Parameters
Comments
Example
Chapter 7
CALibration:STATe:AC <
correction using the calibration constants is enabled or disabled. If
is
, AC correction is disabled. If
OFF
enabled.
AC amplitude correction is only possible when
SOURce:ARBitrary:DAC:SOURce
reference oscillator frequency is 42.94967296 MHz, either internal or
external.
CALibration:STATe:AC
SOURce:ARBitrary:DAC:SOURce VXI
reference oscillator frequency is selected.
The HP E1340A uses an active amplitude correction technique,
boosting the output signal to compensate for filter and other
roll-offs. When waveform generation at high frequencies is
halted either because ARM:LAYer1:COUNt cycles have been
generated, or because an ABORt command was executed, filter
loss disappears, and the output level may exceed the level
specified by SOURce:VOLTage:LEVel:IMMediate:AMPLitude. It
is recommended that AC amplitude correction be used only for
continuous output.
Parameter
Parameter
Name
Type
state
boolean
Executable when initiated: Query form only
Coupled command: Yes
Related commands:
*RST Condition:
Disabling AC amplitude correction
CAL:STAT:AC OFF
state
>
specifies whether AC amplitude
is
STATe
ON
is set to
INTerna
must be set to
OFF
is set or when a different external
Range of
Values
OFF | 0 | ON | 1
SOURce:ROSCillator:FREQuency:EXTernal
CALibration:STATe:AC ON
Disable amplitude correction
HP E1340A Command Reference
CALibration:STATe:AC
STATe
, AC correction is
l and the current
when
Default
Units
none
145