[:STARt]:LAYer2
:SOURce
Parameters
Comments
Example
Chapter 7
ARM[:STARt]:LAYer2:SOURce <
waveform output. The available sources are:
EXTernal:
The HP E1340A's front panel "Aux In" BNC connector.
Immediate arming. An arm is internally generated two
IMMediate:
to three reference oscillator cycles after the start trigger sequence
enters the wait-for-arm state.
Parameter
Parameter
Name
Type
source
discrete
The "Aux In" BNC is a multiple-use input: for FSK control, for
waveform hopping control, as a reference oscillator source, for
reference oscillator gating, or as a start arm source. Only one of
these uses may be active at any time.
If
ARM:STARt:LAYer2:SOURce IMMediate
ARM:STARt:LAYer2:COUNt
ARM:STARt:LAYer2:SOURce EXTernal
ARM:STARt:LAYer2:COUNt
The
ARM:STARt:LAYer2:SLOPe
for the front panel "Aux In" BNC when used as the start arm source.
Executable when initiated: Query form only
Coupled command: Yes
Related Commands:
ARM:STARt:LAYer2:SLOPe
*RST Condition:
Setting the start arm source
ARM:LAY2:SOUR EXT;COUN INF
ARM[:STARt]:LAYer2 :SOURce
source
>
selects the source that will start
Range of
Values
EXTernal | IMMediate
must be set to 1. If
is set,
must be set to
command selects the active edge
ARM:STARt:LAYer2:COUNt
ARM:STARt:LAYer2:SOURce IMMediate
Start arm source is front panel
"Aux In" BNC, and count is infinity
HP E1340A Command Reference
Default
Units
none
is set, then
.
INFinity
,