[SOURce:]FUNCtion

Subsystem Syntax

[:SHAPe]

Note
Parameters
Comments
Chapter 7
The
[SOURce:]FUNCtion
(arbitrary, sinusoid, etc.) the HP E1340A generates. For arbitrary waveform
generation, the subsystem controls which of the seven possible segment
sequences are selected.

[SOURce:]

FUNCtion

shape
[:SHAPe] <
:USER <
name
[SOURce:]FUNCtion[:SHAPe] <
HP E1340A generates, shown as follows:
DC:
Generates a DC output voltage.
Generates a stepped ramp. The
RAMP:
controls the polarity.
SINusoid:
Generates a sinusoidal voltage.
Generates a square wave.
SQUare:
TRIangle:
Generates a stepped triangle wave.
Generates an arbitrary waveform. The
USER:
SOURce:FUNCtion:USER
to be generated.
Selecting a shape other than DC or USER causes the voltage data contained
in waveform segment "D" to be overwritten by the voltage data for the
selected waveform shape. Segment "D" must be reloaded before it can be
output as part of a segment sequence.
Parameter
Parameter
Name
Type
shape
discrete
For the
DC
function: The voltage level is specified by
SOURce:VOLTage:LEVel:IMMediate:AMPLitude
[SOURce:]FUNCtion[:SHAPe]
subsystem controls which waveform shape
>
>
shape
selects what waveform shape the
>
SOURce:RAMP
command selects the segment sequence
Range of
Values
DC | RAMP | SINusoid |
SQUare | TRIangle | USER
HP E1340A Command Reference
subsystem
Default
Units
none
.