ABORt

ABORt
Subsystem Syntax
Comments
Example
138
HP E1340A Command Reference
The
ABORt
command places the trigger subsystem in the idle state,
regardless of any other settings. The command halts waveform generation,
but keeps the output voltage at the value generated when
executed. Only another
output.
ABORt
ABORt
does not affect any other settings of the HP E1340A.
The Pending Operation Flag set true by the
command will be set false as a consequence of entering the trigger
idle state. Subsequent
therefore complete immediately.
Executable when initiated: Yes
Coupled command: No
Related Commands:
*RST Condition:
state, as if executing an
Aborting a waveform
ABOR
INITiate:IMMediate
command will restart waveform
[no query]
*OPC
,
*OPC?
, and
,
,
*OPC
*OPC?
*WAI
places the HP E1340A in the trigger idle
*RST
ABORt
command.
Place HP E1340A in idle state
was
ABORt
INITiate:IMMediate
*WAI
commands will
,
INITiate:IMMediate
Chapter 7