Chapter 3
6. Store the First Waveform Segment as Voltages -
[SOURce:]LIST[:SEGMent]:VOLTage <
This command stores the points of the " A" waveform segment into
the AFG's segment memory. These points are sent to the AFG as
volts which are the output voltage points that constitutes the
waveform segment.
7. Select the Second Waveform Segment -
[SOURce:]LIST[:SEGMent]:SELect B
Use
B
to select the second waveform segment.
8. Store the Second Waveform Segment as Voltages -
[SOURce:]LIST[:SEGMent]:VOLTage <
This command stores the points of the
AFG's segment memory. These points are sent to the AFG as volts
which are the output voltage points that constitutes the waveform
segment.
9. Select the User Name -
[SOURce:]FUNCtion:USER AB
This command selects the waveform segments to be output
(i.e.,
).
AB
10. Initiate the Waveform -
This command generates an immediate output with the arm source
set to
IMMediate
. Refer to Chapter 5 for triggering information.
Generating Arbitrary Waveforms with the HP E1340A
voltage_list
voltage_list
B
waveform segment into the
INITiate[:IMMediate]
>
>