Figure 5-5. Symmetry Adjustment
c.
Measure t1 and adjust A13R22, LOWER
FREQ. SYM., to made t2 = t1. Ref. Figure
5-5.
If A13R22 does not have enough
range change A13Q20.
5-31.
Upper
Frequency
(A13R23).
Lower Frequency Symmetry must be
set before this adjustment is made.
a.
With the same setup as used for the Lower
Symmetry Adjust, select X100 RANGE and
adjust A13R23 to make t2 equal to t1. The
symmetry error should be < 1%.
b.
Check the symmetry with the dial set to 3
and then again with the dial set to 10. The
symmetry error at both dial settings should
be <1%.
% Symmetry error = t1 -t2 X 100
5-32.
CURRENT SOURCE ADJUST (A13R24).
This adjustment interacts with the
Frequency
Symmetry
(A13R22 and A13R23): perform the
following adjustment only if the
frequency is not within specified
accuracy(Table 1-1) on the X10 orX.1
RANGE.
a.
Test
equipment
Counter (-hp- Model 5245L).
b.
Connect
Frequency Counter, and set 3300A controls
as follows:
Symmetry
NOTE
t1 -t2
NOTE
Adjustments
required:
Frequency
CHANNEL
A
OUTPUT
Figure 5-6. DC Output Level Adjust Test Setup
Adjust.
5-33.
5-34.
5-35.
5-36.
to
5-5
RANGE switch ........................... X10
FREQUENCY dial ...................... 10
CHANNEL A function ................. SINE
c.
Output frequency should be 100 Hz, *1
minor division on FREQUENCY dial.
d.
Position RANGE switch to X.
measure output frequency (1 Hz *1 minor
division on FREQUENCY dial.
NOTE
Repeat Frequency Symmetry Adjust
outlined in paragraph 5-29 if A13R24
is adjusted in the following step.
e.
If the frequency is not within specifications
given in step c or d, adjust A13R24
CURRENT SOURCE ADJ for optimum
indication on both X10 and X. I RANGE.

DIAL ADJUSTMENT.

a.
Test
equipment
Counter (-hp- Model 5245L).
b.
Connect
CHANNEL
Frequency Counter, and set 3300A control
as follows:
RANGE switch ........................... X10
CHANNEL A function................. SINE
c.
Loosen dial from hub and adjust the
frequency of 3300A to exactly 100 cps by
rotating the hub. Set the dial to read "1"
and tighten the dial to the hub. Recheck
the frequency.
DIAL CALUBRATE.
FREQUENCY
CALIBRATION
(A13R17).
a.
With same setup as used for 5-33, turn
FREQUENCY dial to "10".
b.
Adjust A13RI7 FREQ CAL ADJ for output
frequency of 1 kHz.
X1K RANGE ADJUST (A13C19).
a.
With same setup as used for 5-33, set
RANGE switch to X1K and FREQUENCY
dial to "10".
Section V
1 and
required:
Frequency
A
OUTPUT
to
ADJUST.