Dc Power Specification Summary; Ac Power (45 Hz To 65 Hz) Specification Summary, Pf=1 - Fluke 5520A Service Manual

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

1-18. DC Power Specification Summary

Voltage Range
90 days
33 mV to 1020 V
1 year
33 mV to 1020 V
[1] To determine dc power uncertainty with more precision, see the individual "AC Voltage Specifications," "AC
Current Specifications," and "Calculating Power Uncertainty."
[2] Add 0.02% unless a settling time of 30 seconds is allowed for output currents
two current ranges within 30 seconds of an output current

1-19. AC Power (45 Hz to 65 Hz) Specification Summary, PF=1

Voltage Range
90 days
33 to 329.999 mV
330 mV to 1020 V
1 year
33 to 329.999 mV
330 mV to 1020 V
Voltage Range
90 days
33 to 329.999 mV
330 mV to 1020 V
1 year
33 to 329.999 mV
330 mV to 1020 V
[1] To determine ac power uncertainty with more precision, see the individual "DC Voltage Specifications" and "DC
Current Specifications" and "Calculating Power Uncertainty."
[2] Add 0.02% unless a settling time of 30 seconds is allowed for output currents
two
current ranges within 30 seconds of an output current
Current Range
0.33 mA to
329.99 mA
Absolute Uncertainty, tcal
0.021
0.023
10 A.
3.3 mA to
9 mA to
8.999 mA
32.999 mA
Absolute Uncertainty, tcal
0.13
0.09
0.11
0.07
0.14
0.10
0.12
0.08
Current Range [2]
0.33 A to
0.9 A to
0.8999 A
2.1999 A
Absolute Uncertainty, tcal
0.12
0.10
0.10
0.08
0.13
0.11
0.11
0.09
10 A.
Introduction and Specifications
0.33 A to
2.9999 A
5 C,
(% of watts output) [1]
0.019 [2]
0.022 [2]
10 A or for currents on the highest
Current Range
33 mA to
89.99 mA
5 C,
(% of watts output) [1]
0.13
0.11
0.14
0.12
2.2 A to
4.4999 A
5 C,
(% of watts output) [1]
0.12
0.11
0.13
0.12
10 A or for currents on the highest
Specifications
3 A to
20.5 A
0.06 [2]
0.07 [2]
90 mA
to 329.99 mA
0.09
0.07
0.10
0.08
4.5 A to
20.5 A
0.10
0.09
0.11
0.10
1-19
1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents