3-24. Verifying AC Voltage (AUX Output)

Verify that the 5520A performance is within the limits in Table 3-25, using the same
equipment and techniques specified previously for calibration.
Range
329.999 mV
329.999 mV
329.999 mV
329.999 mV
329.999 mV
329.999 mV
329.999 mV
329.999 mV
329.999 mV
329.999 mV
329.999 mV
329.999 mV
3.29999 V
3.29999 V
3.29999 V
3.29999 V
3.29999 V
3.29999 V
3.29999 V
5.00000 V
5.00000 V
5.00000 V
5.00000 V
5.00000 V
5.00000 V
Note: set the NORMAL output to 300 mV.
3-42
Table 3-25. Verification Tests for AC Voltage (AUX Output)
Output, AUX
Frequency
(Note)
10.000 mV
45 Hz
10.000 mV
1 kHz
10.000 mV
5 kHz
10.000 mV
10 kHz
10.000 mV
30 kHz
300.000 mV
9.5 Hz
300.000 mV
10 Hz
300.000 mV
45 Hz
300.000 mV
1 kHz
300.000 mV
5 kHz
300.000 mV
10 kHz
300.000 mV
30 kHz
3.00000 V
9.5 Hz
3.00000 V
10 Hz
3.00000 V
45 Hz
3.00000 V
1 kHz
3.00000 V
5 kHz
3.00000 V
10 kHz
3.00000 V
30 kHz
5.00000 V
9.5 Hz
5.00000 V
10 Hz
5.00000 V
45 Hz
5.00000 V
1 kHz
5.00000 V
5 kHz
5.00000 V
10 kHz
Lower Limit
Upper Limit
9.622 mV
10.378 mV
9.622 mV
10.378 mV
9.535 mV
10.465 mV
9.520 mV
10.480 mV
8.700 mV
11.300 mV
283.350 mV
316.650 mV
299.180 mV
300.820 mV
299.390 mV
300.610 mV
299.390 mV
300.610 mV
299.100 mV
300.900 mV
298.650 mV
301.350 mV
287.100 mV
312.900 mV
2.825 V
3.175 V
2.99505 V
3.00495 V
2.99745 V
3.00255 V
2.99745 V
3.00255 V
2.99410 V
3.00590 V
2.98960 V
3.01040 V
2.87720 V
3.12280 V
4.72500 V
5.27500 V
4.99205 V
5.00795 V
4.99605 V
5.00395 V
4.99605 V
5.00395 V
4.99110 V
5.00890 V
4.98360 V
5.01640 V