Fluke 5520A Service Manual page 192

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

5520A
Service Manual
Calibrator
Mainframe
Wave
Type
square
square
square
square
square
square
square
square
square
square
square
square
square
square
square
square
square
square
sine
sine
sine
sine
sine
sine
sine
triangle
triangle
triangle
triangle
triangle
triangle
triangle
6-52
Table 6-36. Wave Generator Verification at 50
Calibrator
5790A
Mainframe
Reading
Conversion
output
(V rms)
(10 kHz)
1.8 mV
2.0000
6.4 mV
2.0000
10.9 mV
2.0000
11.0 mV
2.0000
28.0 mV
2.0000
44.9 mV
2.0000
45 mV
2.0000
78 mV
2.0000
109 mV
2.0000
110 mV
2.0000
280 mV
2.0000
449 mV
2.0000
450 mV
2.0000
780 mV
2.0000
1.09 V
2.0000
1.10 V
2.0000
1.80 V
2.0000
2.50 V
2.0000
1.8 mV
2.8284
10.9 mV
2.8284
44.9 mV
2.8284
109 mV
2.8284
449 mV
2.8284
1.09 V
2.8284
2.50 V
2.8284
1.8 mV
3.4641
10.9 mV
3.4641
44.9 mV
3.4641
109 mV
3.4641
449 mV
3.4641
1.09 V
3.4641
2.50 V
3.4641
5790A Rdg x
Conversion
Factor
Factor (V p-p)
V p-p value x
Tolerance
(V p-p)
correction
0.000154 V
0.000292 V
0.000427 V
0.00043 V
0.00094 V
0.001447 V
0.00145 V
0.00244 V
0.00337 V
0.0034 V
0.0085 V
0.01357 V
0.0136 V
0.0235 V
0.0328 V
0.0331 V
0.0541 V
0.0751 V
0.000154 V
0.000427 V
0.001447 V
0.00337 V
0.01357 V
0.0328 V
0.0751 V
0.000154 V
0.000427 V
0.001447 V
0.00337 V
0.01357 V
0.0328 V
0.0751 V

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents