6-60. Leveled Sine Wave Flatness Verification

Leveled Sine Wave flatness verification is divided into two frequency bands: 50 kHz to
10 MHz (low frequency) and > 10 MHz to 600 MHz (high frequency). The equipment
setups are different for each band. Leveled Sine Wave flatness is measured relative to 50
kHz. This is determined directly in the low frequency band. The high frequency band
requires a "transfer" measurement be made at 10 MHz to calculate a flatness relative to
50 kHz.

6-61. Equipment Setup for Low Frequency Flatness

All low frequency flatness procedures use the following equipment.
Connect the Calibrator Mainframe SCOPE connector to the 5790A WIDEBAND input
with the BNC(f) to Type N(m) adapter as shown in Figure 6-10. Set the 5790A to
AUTORANGE, digital filter mode to FAST, restart fine, and Hi Res on.
5790A
AC MEASUREMENT
STANDARD
INPUT 1
1000V RMS MAX
SHELL FLOATING
SHUNT
3V RMS MAX
WIDEBAND
7V RMS MAX
SHELL FLOATING
GROUND
10V PK
MAX
Figure 6-10. Connecting the Calibrator Mainframe to the 5790A AC Measurement Standard

6-62. Equipment Setup for High Frequency Flatness

All high frequency flatness procedures use the following equipment.
When high frequencies at voltages below 63 mV p-p are verified, use the
8481D Power Sensor. Otherwise, use the 8482A Power Sensor.
6-44
5790A/03 AC Measurement Standard with Wideband option
BNC(f) to Type N(m) adapter
BNC cable supplied with the SC600
INPUT 2
1000V RMS MAX
INPUT1
INPUT1
INPUT1
SHUNT
INPUT1
HI
.
VIEW
2.2 mV
22 mV
220 mV
2.2 V
22 V
220 mV
REF
6
7
8
9
+/-
DELETE
7 mV
70 mV
700 mV
7V
70V
700V
LO
0
1
2
3
4
5
CLEAR
10V PEAK
MAX
2.2 mV
1kV
AUTO
MAN
ENTER
GUARD
Hewlett-Packard 437B Power Meter
Hewlett-Packard 8482A and 8481D Power Sensors
BNC(f) to Type N(f) adapter
BNC cable supplied with the Calibrator Mainframe
5520A
CALIBRATOR
NORMAL
AUX
SCOPE
V, , ,RTD
A, -SENSE, AUX V
OUT
HI
1000V
20V
RMS
1V PK
RMS
UTIL
SPEC
MAX
MAX
MAX
MENUS
LO
TRIG
20V
RMS
MAX
20A
SHELLS
GUARD
NOT
GROUNDED
POWER
I
20V PK MAX
TC
20V PK MAX
O
Note
PREV
STBY
OPR
EARTH
SCOPE
BOOST
MENU
dBm
sec
EDIT
7
8
9
SETUP
RESET
150V
m
V
Hz
FIELD
PK
MAX
n
W
¡F
NEW
4
5
6
k
A
¡C
CE
REF
p
MEAS
TRIG
20V
1
2
3
M
F
PK
TC
OUT
MAX
MULT
DIV
/ +
0
SHIFT
ENTER
x
POWER
I
O
yg034f.eps