6-27. Equipment Required for Calibration and Verification

Table 6-15 lists the equipment, recommended models, and minimum specifications
required for each calibration and verification procedure.
Table 6-15. SC600 Calibration and Verification Equipment
Wave Generator and Edge Amplitude Calibration, AC Voltage and TD Pulser Verification
Instrument
Digital
HP 3458A
Multimeter
Adapter
Pomona #1269
Termination
BNC Cable
(supplied with SC600)
High-
Tektronix 11801 with
Frequency
Tektronix SD-22/26
Digital Storage
sampling head, or
Oscilloscope
Tektronix TDS 820 with
8 GHz bandwidth
Attenuator
Weinschel 9-10 (SMA)
or Weinschel 18W-10
or equivalent
Adapter
BNC Cable
(supplied with SC600)
Leveled Sine Wave Amplitude Calibration and Verification
AC
Fluke 5790A
Measurement
Standard
Adapter
Pomona #1269
Termination
BNC Cable
(supplied with SC600)
DC and AC Voltage Calibration and Verification, DC Voltage Verification
Digital
HP 3458A
Multimeter
Adapter
Pomona #1269
Termination
BNC Cable
(supplied with SC600)
Model
Voltage
Edge
BNC(f) to Double Banana Plug
Feedthrough 50
Calibration and AC Voltage Verification)
Edge Rise Time and Aberrations Verification
Frequency
Resolution
10 dB, 3.5 mm (m/f)
BNC(f) to 3.5 mm(m)
Range
Frequency
BNC(f) to Double Banana Plug
Feedthrough 50
BNC(f) to Double Banana Plug
Feedthrough 50

Equipment Required for Calibration and Verification

Minimum Use Specifications
1.8 mV to 130 V p-p Uncertainty: 0.06%
4.5 mV to 2.75 V p-p Uncertainty: 0.06%
1% (used with Edge Amplitude
12.5 GHz
4.5 mV to 2.75 V
5 mV p-p to 5.5 V p-p
50 kHz
1%.
1%.
SC600 Option
6
6-15