Ac Current (Non-Sinewave) Specifications - Fluke 5520A Service Manual

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

1-32. AC Current (Non-Sine Wave) Specifications
Triangle Wave &
Truncated Sine
Wave Range
p-p
0.047 mA
0.01 Hz o 10 Hz
to 0.92999 mA [1]
10 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 10 kHz
0.93 mA to
0.01 Hz to 10 Hz
9.29999 mA [1]
10 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 10 kHz
9.3 mA to
0.01 Hz to 10 Hz
92.9999 mA [1]
10 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 10 kHz
93 mA to
0.01 Hz to 10 Hz
929.999 mA [1]
10 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 10 kHz
0.93 A to
10 Hz to 45 Hz
8.49999 A
45 Hz to 1kHz
1 kHz to 10 kHz
8.5 A to 57 A [2]
45 Hz to 500 Hz
500 Hz to 1 kHz
[1] Frequency limited to 1 kHz with LCOMP on.
[2] Frequency limited to 440 Hz with LCOMP on
Frequency
1-Year Absolute Uncertainty
(% of output
5.0
0.25
0.25
10
5.0
0.25
0.25
10
5.0
0.25
0.25
10
5.0
0.25
0.25
10
0.5
0.5
10
0.5
1.0
Introduction and Specifications
tcal
5 C
% of range)
0.5
0.5
0.25
2
0.5
0.5
0.25
2
0.5
0.5
0.25
2
0.5
0.5
0.5
2
1.0
0.5
2
0.5
1.0
Additional Specifications
Max
Current
Resolution
Two digits
Six digits
Two digits
Six digits
Two digits
Six digits
Two digits
Six digits
Six digits
1