Ac Voltage (Non-Sinewave) Specifications - Fluke 5520A Service Manual

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

5520A
Service Manual
1-27. AC Voltage (Non-Sine Wave) Specifications
Triangle Wave
&
Truncated
Sine
Range, p-p [1]
2.9 mV
to 93 V
93 mV
to 14 V
[1] To convert p-p to rms for triangle wave, multiply the p-p value by 0.2886751. To convert p-p to rms for
truncated sine wave, multiply the p-p value by 0.2165063.
[2] Uncertainty is stated in p-p. Amplitude is verified using an rms-responding DMM.
[3] Uncertainty for Truncated Sine outputs is typical over this frequency band.
Square
Wave Range
(p-p) [1]
2.9 mV
to
66 V
66 mV
to
14V
[1]
To convert p-p to rms for square wave, multiply the p-p value by 0.5.
[2]
Uncertainty is stated in p-p. Amplitude is verified using an rms-responding DMM.
[3]
Limited to 1 kHz for Auxiliary outputs
1-26
Frequency
Normal Channel (Single Output Mode)
0.01 Hz to 10 Hz
10 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 20 kHz
20 kHz to 100 kHz
[3]
Auxiliary Output (Dual Output Mode)
0.01 Hz to 10 Hz
10 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 10 kHz
1-Year Absolute Uncertainty,
Frequency
(% of output
Normal Channel (Single Output Mode)
0.01 Hz to 10 Hz
10 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 20 kHz
20 kHz to 100 kHz
Auxiliary Output (Dual Output Mode)
0.01 Hz to 10 Hz
10 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 10 kHz [3]
1-Year Absolute Uncertainty,
tcal
5 C,
(% of output
% of range) [2]
5.0
0.5
0.25
0.5
0.25
0.25
0.5
0.25
5.0
0.5
5.0
0.5
0.25
0.5
0.25
0.25
5.0
0.5
tcal
5 C
% of range) [2]
5.0
0.5
0.25
0.5
0.25
0.25
0.5
0.25
5.0
0.5
5.0
0.5
0.25
0.5
0.25
0.25
5.0
0.5
6.6 V p-p.
Max Voltage
Resolution
Two digits on
each range
Six digits on
each range
Two digits on
each range
Six digits on
each range
Max Voltage
Resolution
Two digits on each range
Six digits on each range
Two digits on each range
Six digits on each range

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents