DC Volts Calibration (NORMAL Output)

The equipment listed in Table 3-2 is required for calibration of the dc volts function. (The
equipment is also listed in the consolidated table, Table 3-1.)
Quan.
1
Fluke
1
Hewlett Packard
1
Fluke
1
Keithley
Proceed as follows to calibrate the dc voltage function:
1. On the HP 3458A, perform the ACAL (autocal) ALL and MATH NULL functions as
2. Verify that the UUT (Unit Under Test) is in Standby.
3. Start 5520A calibration as described under the previous heading.
4. Perform an internal DC Zeros Calibration as prompted.
5. Connect the test equipment as shown in Figure 3-1.
6. Measure and enter the values into the UUT for steps 1 through 6 in Table 3-3 as
7. Verify that the UUT is in Standby.
8. Connect the DMM and Reference Divider to the UUT as shown in Figure 3-2.
9. For voltages 30 Vdc and above, see the next section.
3-6
Table 3-2. Test Equipment Required for Calibrating DC Volts
Manufacturer
5500A/LEADS
3458A with -002 Option
752A
155
described in the HP 3458A user documentation.
prompted. You will need to disconnect and reconnect the DMM as prompted during
these steps.
Model
Test lead set
DMM
Reference Divider
Null Detector
Equipment