Additional Specifications; Frequency Specifications - Fluke 5520A Service Manual

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

1-23. Additional Specifications

The following paragraphs provide additional specifications for the 5520A Calibrator ac
voltage and ac current functions. These specifications are valid after allowing a warm-up
period of 30 minutes, or twice the time the 5520A has been turned off. All extended
range specifications are based on performing the internal zero-cal function at weekly
intervals, or when the ambient temperature changes by more than 5 C.

1-24. Frequency Specifications

Frequency
Range
0.01 Hz to 119.99 Hz
120.0 Hz to 1199.9 Hz
1.200 kHz to 11.999 kHz
12.00 kHz to 119.99 kHz
120.0 kHz to 1199.9 kHz
1.200 MHz to 2.000 MHz
[1] With REF CLK set to ext, the frequency uncertainty of the 5520A is the uncertainty of the external
10 MHz clock
5 Hz. The amplitude of the 10 MHz external reference clock signal should be
between 1 V and 5 V p-p.
1-Year Absolute Uncertainty,
Resolution
0.01 Hz
0.1 Hz
1.0 Hz
10 Hz
100 Hz
1 kHz
Introduction and Specifications
Additional Specifications
tcal
5 C
2.5 ppm
5 Hz [1]
1
Jitter
100 nS
1-23