Voltage
Input Frequency
100 Hz
1 kHz
5 kHz
10 kHz
For all measurements, the HP 3458A is in DCV, manual ranging, with external trigger
enabled. A convenient method to make these measurements from the HP 3458A's front
panel is to program these settings into several of the user defined keys on its front panel.
For example, to make topline measurements at 1 kHz, you would set the DMM to "NPLC
.01; DELAY .0007; TRIG EXT". To find the average of multiple readings, you can
program one of the keys to "MATH OFF; MATH STAT" and then use the "RMATH
MEAN" function to recall the average or mean value.
For this application, if making measurements of a signal > 1 kHz, the HP
3458A has been known to have .05% to .1% peaking in the 100 mV range.
For these signals, lock the HP 3458A to the 1 V range.
HP 3458A (Front)
BNC(F) to
Double Banana
Adapter
Figure 6-2. Equipment Setup for SC600 Voltage Square Wave Measurements
Calibration and Verification of Square Wave Voltage Functions
Table 6-16. Voltage HP3458A Settings
NPLC
.1
.01
.002
.001
Note
SC600 Cable
50
Feedthrough
Termination
HP 3458A (Rear)
HP 3458A Settings
DELAY (topline)
.007 s
.0007 s
.00014
.00007
SC600 Option
DELAY (baseline)
.012 s
.0012 s
.00024
.00012
5520A-SC600
5520A
CALIBRATOR
NORMAL
AUX
SCOPE
V, , ,RTD
A, -SENSE, AUX V
OUT
HI
1000V
20V
1V PK
RMS
RMS
MAX
MAX
MAX
150V
LO
PK
MAX
TRIG
20V
RMS
MAX
20V
SHELLS
PK
20A
GUARD
NOT
MAX
GROUNDED
20V PK MAX
TC
20V PK MAX
ygo54f.eps
6
6-19