Dc Power Specification Summary; Ac Power (45 Hz To 65 Hz) Specification Summary, Pf=1 - Fluke 5520A Getting Started

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

5520A
Getting Started

DC Power Specification Summary

90
33 mV to 1020 V
days
1 year
33 mV to 1020 V
[1] To determine dc power uncertainty with more precision, see the individual "AC Voltage Specifications," "AC
Current Specifications," and "Calculating Power Uncertainty."
[2] Add 0.02% unless a settling time of 30 seconds is allowed for output currents > 10 A or for currents on the highest
two current ranges within 30 seconds of an output current > 10 A.

AC Power (45 Hz to 65 Hz) Specification Summary, PF=1

90
33 to 329.999 mV
days
330 mV to 1020 V
1 year
33 to 329.999 mV
330 mV to 1020 V
90
days
1 year
[1]
To determine ac power uncertainty with more precision, see the individual "DC Voltage Specifications" and "DC
Current Specifications" and "Calculating Power Uncertainty."
Add 0.02% unless a settling time of 30 seconds is allowed for output currents > 10A or for currents on the highest
[2]
two current ranges within 30 seconds of an output current > 10A.
20
0.33 mA to
329.99 mA
Voltage Range
3.3 mA to
8.999 mA
Voltage Range
Voltage Range
33 to 329.999 mV
330 mV to 1020 V
33 to 329.999 mV
330 mV to 1020 V
Current Range
0.33 A to
2.9999 A
Absolute Uncertainty, tcal ± 5 °C, ± (% of watts output) [1]
0.021
0.019 [2]
0.023
0.022 [2]
Current Range
9 mA to
32.999 mA
Absolute Uncertainty, tcal ± 5 °C, ± (% of watts output) [1]
0.13
0.09
0.11
0.07
0.14
0.10
0.12
0.08
Current Range [2]
0.33 A to
0.9 A to
0.8999 A
2.1999 A
Absolute Uncertainty, tcal ± 5 °C, ± (% of watts output) [1]
0.12
0.10
0.10
0.08
0.13
0.11
0.11
0.09
3 A to
20.5 A
0.06 [2]
0.07 [2]
33 mA to
90 mA
89.99 mA
to 329.99 mA
0.13
0.09
0.11
0.07
0.14
0.10
0.12
0.08
2.2 A to
4.5 A to
4.4999 A
20.5 A
0.12
0.10
0.11
0.09
0.13
0.11
0.12
0.10

Advertisement

Table of Contents
loading

Table of Contents