AC Current (Non-Sine Wave) Specifications (cont)
Square Wave
Range
p-p
0.047 mA to
0.65999 mA [1]
0.66 mA to
6.59999 mA [1]
6.6 mA to
65.9999 mA [1]
66 mA to
659.999 mA [1]
0.66 A to
5.99999 A [2]
6 A to 41 A [2]
[1] Frequency limited to 1 kHz with LCOMP on.
[2] Frequency limited to 440 Hz with LCOMP on.
1-30
Frequency
1-Year Absolute Uncertainty,
(% of output
0.01 Hz to 10 Hz
10 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 10 kHz
0.01 Hz to 10 Hz
10 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 10 kHz
0.01 Hz to 10 Hz
10 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 10 kHz
0.01 Hz to 10 Hz
10 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 10 kHz
10 Hz to 45 Hz
45 Hz to 1 kHz
1 kHz to 10 kHz
45 Hz to 500 Hz
500 Hz to 1 kHz
tcal
5 C,
% of range)
5.0
0.5
0.25
0.5
0.25
0.25
10
2
5.0
0.5
0.25 0.5
0.25
0.25
10
2
5.0
0.5
0.25
0.5
0.25
0.25
10
2
5.0
0.5
0.25
0.5
0.25
0.5
10
2
0.5
1.0
0.5
0.5
10
2
0.5
0.5
1.0
1.0
Max
Current
Resolution
Two digits
Six digits
Two digits
Six digits
Two digits
Six digits
Two digits
Six digits