3-19. Verifying DC Volts (NORMAL Output)

Verify that the 5520A performance is within the limits in Table 3-19, using the same
equipment and techniques specified previously for calibration.
Table 3-19. Verification Tests for DC Voltage (NORMAL Output)
Range
329.9999 mV
329.9999 mV
329.9999 mV
3.299999 V
3.299999 V
3.299999 V
3.299999 V
3.299999 V
32.99999 V
32.99999 V
32.99999 V
32.99999 V
32.99999 V
329.9999 V
329.9999 V
329.9999 V
329.9999 V
1000.000 V
1000.000 V
1000.000 V
1000.000 V
1000.000 V
1000.000 V
Output
0.0000 mV
329.0000 mV
-329.0000 mV
0.000000 V
1.000000 V
-1.000000 V
3.290000 V
-3.290000 V
0.00000 V
10.00000 V
-10.00000 V
32.90000 V
-32.90000 V
50.0000 V
329.0000 V
-50.0000 V
-329.0000 V
334.000 V
900.000 V
1020.000 V
-334.000 V
-900.000 V
-1020.000 V
Calibration and Verification
Performance Verification Tests
Lower Limit
-0.0010 mV
328.9941 mV
-329.0059 mV
-0.000002 V
0.999989 V
-1.000011 V
3.289968 V
-3.290032 V
-0.00002 V
9.99988 V
-10.00012 V
32.89965 V
-32.90035V
49.9991 V
328.9949 V
-50.0009 V
-329.0051 V
333.993 V
899.985 V
1019.983 V
-334.007 V
-900.015 V
-1020.017 V
Upper Limit
0.0010 mV
329.0059 mV
-328.9941 mV
0.000002 V
1.000011 V
-0.999989 V
3.290032 V
-3.289968 V
0.00002 V
10.00012 V
-9.99989 V
32.90035 V
-32.89965 V
50.0009 V
329.0051 V
-49.9991 V
-328.9949 V
334.007 V
900.015 V
1020.017 V
-333.993 V
-899.985 V
-1019.983 V
3
3-35