Fluke 5520A Service Manual page 85

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

CAL_ABORT
Description:
Instruct 5520A to abort calibration procedure after present step
Example:
CAL_ABORT
CAL_BACKUP
Description:
Skip to next entry point in calibration procedure.
CAL_DATE?
Description:
Return a calibration date associated with stored calibration constants.
The date is returned with the same format as the CLOCK command.
Parameter:
Which date: MAIN, ZERO, OHMSZERO, SCOPE
Response:
The date
CAL_DAYS?
Description:
Return the number of days and hours since the last calibration constants
were stored.
Parameter:
Which date: MAIN, ZERO, OHMSZERO, SCOPE
Response:
1. (Integer) Days
2. (Integer) Hours
CAL_FACT
Description:
Set the procedure "fault action" flag. Procedures refer to both calibration
and diagnostic procedures. This command is more useful for diagnostics
than calibration.
Parameter:
(Character) CONT to continue on faults or ABORT to abort on faults
Example:
CAL_FACT ABORT (this is the default)
CAL_FACT?
Description:
Get the procedure "fault action" flag
Response:
(Character) CONT or ABORT
Example:
ABORT
CAL_FAULT?
Description:
Get information about calibration error (if one occurred)
Response:
1. error number (use EXPLAIN? command to interpret)
2. Name of step where error occurred
Calibration and Verification
Calibration Remote Commands
3
3-29