Table 6-15. SC600 Calibration and Verification Equipment (cont.)
Leveled Sine Wave Flatness (High Frequency) Calibration and Verification
Instrument
Power Meter
Hewlett-Packard 437B
Power Sensor
Hewlett-Packard 8482A
Power Sensor
Hewlett-Packard 8481D
30 dB
Hewlett-Packard
Reference
11708A
Attenuator
(supplied with HP
8481D)
Adapter
Hewlett-Packard
PN 1250-1474
BNC Cable
(supplied with SC600)
Leveled Sine Wave Frequency, Time Marker Verification
Frequency
PM 6680 with option
Counter
(PM 9621, PM 9624, or
PM 9625) and (PM
9690 or PM 9691)
Adapter
Pomona #3288
BNC Cable
(supplied with SC600)
AC
Fluke 5790A
Measurement
Standard
Adapter
Pomona #1269
Termination
BNC Cable
(supplied with SC600)

6-28. SC600 Calibration Setup

The procedures in this manual have been developed to provide users the ability to
calibrate the SC600 at their own site if they are required to do so. It is strongly
recommended that, if possible, you return your unit to Fluke for calibration and
verification. The Calibrator Mainframe must be fully calibrated prior to performing any
of the SC600 calibration procedures.
The hardware adjustments are intended to be one-time adjustments performed in the
factory, however, adjustment may be required after repair. Hardware adjustments must be
performed prior to calibration. Calibration must be performed after any hardware
adjustments. See "Hardware Adjustments" in this chapter.
The AC Voltage function is dependent on the DC Voltage function. Calibration of the
AC Voltage function is required after the DC Voltage is calibrated.
The Calibrator Mainframe must complete a warm-up period and the SC600 must be
enabled for at least 5 minutes prior to calibration to allow internal components to
Model
Range
Frequency
Range
Frequency
Range
Frequency
Range
Frequency
BNC(f) to Type N(f)
2 ns to 5 s, 50 kHz to 600 MHz: < 0.15 ppm uncertainty
BNC(f) to Type N(m)
Wave Generator Verification
Range
Frequency
BNC(f) to Double Banana
Feedthrough 50
Minimum Use Specifications
-42 to +5.6 dBm
10 - 600 MHz
-20 to +19 dBm
10 - 600 MHz
-42 to -20 dBm
10 - 600 MHz
30 dB
50 MHz
1.8 mV p-p to 55 V p-p
10 Hz to 100 kHz
1%.
SC600 Option

SC600 Calibration Setup

6
6-17