General Specifications - Fluke 5520A Service Manual

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

1-9.

General Specifications

Warmup Time
Settling Time
Standard Interfaces
Temperature Performance
Temperature Coefficient
Relative Humidity [1]
Altitude
Safety
Analog Low Isolation
EMC
Line Power [2]
Power Consumption
Dimensions
Weight (without options)
Absolute Uncertainty Definition
Specification Confidence
Interval
[1] After long periods of storage at high humidity, a drying out period (with the power on) of at least one week may be
required.
[2] For optimal performance at full dual outputs (e.g. 1000 V, 20A) choose a line voltage setting that is ± 7.5% from
nominal.
[3] The DC Current ranges 0 to 1.09999 A and 1.1 A to 2.99999 A are sensitive to storage temperatures above 50 °C.
If the 5520A is stored above 50 °C for greater than 30 minutes, these ranges must be re-calibrated. Otherwise, the
90 day and 1 year uncertainties of these ranges double.
Twice the time since last warmed up, to a maximum of 30 minutes.
Less than 5 seconds for all functions and ranges except as noted.
IEEE-488 (GPIB), RS-232, 5725A Amplifier
Operating: 0 °C to 50 °C
Calibration (tcal): 15 °C to 35 °C
Storage: -20 °C to 70 °C [3]
Temperature Coefficient for temperatures outside tcal + °C is 0.1X/°C of
the 90-day specification (or 1-year, as applicable) per °C.
Operating: <80% to 30 °C, <70% to 40 °C, <40% to 50 °C
Storage: <95%, non-condensing
Operating: 3,050 m (10,000 ft) maximum
Non-operating: 12,200 m (40,000 ft) maximum
Complies with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994;
CAN/CSA-C22.2 No. 1010.1-92
20 V
Designed to comply with FCC Rules Part 15; VFG 243/1991. If used in
areas with Electromagnetic fields of 1 to 3 V/m, resistance outputs have a
floor adder of 0.508 •. Performance not specified above 3 V/m. This
instrument may be susceptible to electro-static discharge (ESD) from direct
contact to the binding posts. Good static aware practices should be
followed when handling this and other pieces of electronic equipment.
Line Voltage (selectable): 100 V, 120 V, 220 V, 240 V
Line Frequency: 47 Hz to 63 Hz
Line Voltage Variation: ±10% about line voltage setting
5500A Calibrator, 300 VA; 5725A Amplifier, 750 VA
5500A Calibrator:
Height: 17.8 cm (7 inches), standard rack increment, plus 1.5 cm (0.6
inch) for feet on bottom of unit;
Width: 43.2 cm (17 inches), standard rack width
Depth: 47.3 cm (18.6 inches) overall 5725A Amplifier:
Height, 13.3 cm (5.25 inches), standard rack increment, plus 1.5 cm (0.6
inch) for feet on bottom of unit;
Width, 43.2 cm (17 inches), standard rack width
Depth, 63.0 cm (24.8 inches) overall.
5500A Calibrator, 22 kg (49 lb); 5725A Amplifier 32 kg (70 pounds)
The 5500A specifications include stability, temperature coefficient, linearity,
line and load regulation, and the traceability of the external standards used
for calibration. You do not need to add anything to determine the total
specification of the 5520A for the temperature range indicated.