1-20. Power and Dual Output Limit Specifications

Frequency
dc
10 Hz to 45 Hz
45 Hz to 65 Hz
65 Hz to 500 Hz
65 Hz to 500 Hz
500 Hz to 1 kHz
1 kHz to 5 kHz
5 kHz to 10 kHz
[1] In dual voltage output mode, voltage is limited to 3.3 V to 500 V in the NORMAL output.
[2] In dual voltage output mode, voltage is limited to 3.3 V to 250 V in the NORMAL output.
Notes
The range of voltages and currents shown in "DC Voltage Specifications," "DC Current Specifications," "AC
Voltage (Sine Wave) Specifications," and "AC Current (Sine Wave) Specifications" are available in the
power and dual output modes (except minimum current for ac power is 0.33 mA). However, only those
limits shown in this table are specified. See "Calculating Power Uncertainty" to determine the uncertainty at
these points.
The phase adjustment range for dual ac outputs is 0 to
outputs is 0.01 .
1-20
Voltages
(NORMAL)
0 to
1020 V
33 mV to 32.9999 V
33 mV to 1000 V
330 mV to 1000 V
3.3 V to 1000 V
330 mV to 1000 V
3.3 V to 1000 V [1]
3.3 V to 1000 V [2]
Currents
0 to
20.5 A
0 to
3.3 mA to 2.99999 A
10 mV to 5 V
3.3 mA to 20.5 A
10 mV to 5 V
33 mA to 2.99999 A
100 mV to 5 V
33 mA to 20.5 A
100 mV to 5 V
33 mA to 20.5 A
100 mV to 5 V
33 mA to 2.99999 A
100 mV to 5 V [1]
33 mA to 329.99 mA
1 V to 5 V [2]
179.99 . The phase resolution for dual ac
Power
Voltages
Factor
(AUX)
(PF)
7 V
0 to 1
0 to 1
0 to 1
0 to 1
1
1
1