Generating A Calibration Report - Fluke 5520A Service Manual

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

RPT_STR
Description: Sets the user report string used for calibration reports. The string is stored
in nonvolatile memory. The CALIBRATION switch must be set to
ENABLE.
Parameter:
String of up to 40 characters
RPT_STR?
Description: Returns the user report string used for calibration reports.
Parameter:
None
Response:
(String) Up to 40 characters
STOP_PR
Description: Terminates printing a calibration report if one was being printed.
Parameter:
None
UNCERT?
Description: Returns specified uncertainties for the present output. If there is no
specification for an output, the uncertainty returned is zero.
Parameter:
1. (Optional) The preferred unit in which to express the primary output
2. (Optional) The preferred unit in which to express the secondary output
Response:
1. (Float) 90 day specified uncertainty of primary output.
2. (Float) 1 year specified uncertainty of primary output.
3. (Character) unit of primary output uncertainty.
4. (Float) 90 day specified uncertainty of secondary output.
5. (Float) 1 year specified uncertainty of secondary output.
6. (Character) unit of secondary output uncertainty.
Example:
With a power output of 1V, 1A, 1kHz:

3-16. Generating a Calibration Report

Three different calibration reports are available from the 5520A, each one either
formatted for printing, or in comma-separated variable format for importation into a
spreadsheet. Using the REPORT SETUP softkey under UTILITY FUNCTS / CAL, you
select lines per page, calibration interval, type of report, format, and which serial port to
use. The specification shown in these reports depends on the interval selected in the
REPORT SETUP menu.
The three types of report are as follows:
"stored," lists output shifts as a result of the most recent stored calibration constants.
uncertainty (default is PCT).
uncertainty (default is same as primary unit).
UNCERT?
Returns 2.00E-02,2.10E-02,PCT,4.60E-02,6.00E-02,PCT
Calibration and Verification

Generating a Calibration Report

3
3-33