Dc Power Summary; Ac Power Summary; Power And Dual Output Limits - Fluke 5080A Service Manual

Hide thumbs Also See for 5080A:
Table of Contents

Advertisement

DC Power Summary

Time
Voltage
90 days
33 mV to 1020 V
1 year
33 mV to 1020 V
[1]
To determine the actual dc power specification, see the individual "DC Voltage Specifications", "DC Current Specifications", and
"Calculating Power Specifications" sections. The actual specification at the operating point will usually be significantly better than
the table value, since the specifications state the minimum performance for the voltages and currents listed.

AC Power Summary

Time
Voltages
33 to 329.999 mV
90 days
330 mV to 1020 V
33 to 329.999 mV
1 year
330 mV to 1020 V
33 to 329.999 mV
90 days
330 mV to 1020 V
33 to 329.999 mV
1 year
330 mV to 1020 V
Notes
To determine the actual ac power specification, see the individual "AC Voltage Specifications", "AC Current Specifications", "Phase
Specifications", and "Calculating Power Specifications" sections. The actual specification at the operating point will usually be
significantly better than the table value, since the specifications state the minimum performance for the voltages and currents listed.

Power and Dual Output Limits

Frequency
Voltages (NORMAL)
DC
45 to 65 Hz
33 mV to 1000 V
330 mV to 1000 V
65 to 500 Hz
3.3 V to 1000 V
500 Hz to 1 kHz
330 mV to 1000 V
Notes
The range of voltages and currents shown in "DC Voltage Specifications", "DC Current Specifications", "AC Voltage Specifications", and
"AC Current Specifications" are available in the power and dual output modes, except that the minimum current for AC power is 0.33 mA
However, only the voltages and currents shown in this table are specified. See "Calculating Power Specifications" to determine the
specification at any points within this table.
The phase adjustment range for dual AC outputs is 0 ° to ±179.9 °. The phase resolution for dual AC outputs is 0.1 degree.
Power and dual output amplitude settling times are typically <9 seconds.
0.33 to 3.2999 mA
Specification, tcal ±5 °C, ±(% of watts output)
0.14
0.15
3.3 to 8.9999 mA
Specification, tcal ±5 °C, 45 to 65 Hz, PF = 1, ±(% of watts output)
0.56
0.50
0.58
0.51
0.33 to 0.8999 A
Specification, tcal ±5 °C, 45 to 65 Hz, PF = 1, ±(% of watts output)
0.57
0.51
0.59
0.52
Currents
0 to ±1020 V
0 to ±20.5 A
3.3 mA to 20.5 A
33 mA to 2.9999 A
33 mA to 20.5 A
33 mA to 20.5 A
Currents
3.3 to 329.99 mA
0.33 to 2.9999 A
0.11
0.21
0.11
0.22
Currents
9 to 32.999 mA
33 to 89.99 mA
0.43
0.56
0.34
0.50
0.45
0.58
0.36
0.51
Currents
0.9 to 2.1999 A
2.2 to 4.499 A
0.43
0.54
0.35
0.47
0.46
0.56
0.37
0.49
Voltages (AUX)
0 to ±7 V
100 mV to 5 V
100 mV to 5 V
100 mV to 5 V
100 mV to 5 V
Calibrator
Specifications
3 to 20.5 A
[1]
0.52
0.54
90 to 329.99 mA
0.43
0.34
0.45
0.36
4.5 to 20.5 A
0.69
0.64
0.72
0.67
Power Factor (PF)
-
0 to 1
0 to 1
0 to 1
1
11