Fluke Corporation - Worldwide Support Center MET/CAL Procedure
=============================================================================
INSTRUMENT:
DATE:
AUTHOR:
REVISION:
ADJUSTMENT THRESHOLD:
NUMBER OF TESTS:
NUMBER OF LINES:
CONFIGURATION:
=============================================================================
STEP
# 10 Sep 98 changed Cal_Info? commands to Out? and checked for 10A -
# needs cal_next to get past display; check for 0 out when ACI is done.
#
1.001
1.002
# Set M[10] to 3mA initially
1.003
# Reset UUT - get it out of calibration mode.
1.004
1.005
1.006
1.007
1.008
1.009
1.010
1.011
1.012
1.013
1.013
1.013
1.013
1.013
1.013
1.013
1.014
# Reset 5790A standard.
1.015
1.016
1.017
1.018
1.019
1.020
1.021
1.022
3-18
Sub Fluke 5520A ACI ADJ
22-Sep-98
Gary Bennett, Metrology Specialist
0.6
70%
1
487
Fluke 5790A
FSC
RANGE NOMINAL
ASK-
R
Q N
HEAD
AC CURRENT ADJUSTMENT
MATH
M[10] = 0.003
IEEE
*CLS;*RST; *OPC?[I]
IEEE
ERR?[I$][GTL]
MATH
MEM1 = FLD(MEM2,1,",")
JMPT
IEEE
CAL_SW?[I][GTL]
MEME
JMPZ
1.012
JMP
1.015
HEAD
WARNING! CALIBRATION SWITCH IS NOT ENABLED.
DISP
The UUT CALIBRATION
DISP
DISP
The switch MUST be in ENABLE to store the
DISP
new calibration constants.
DISP
DISP
Select ENABLE, then press "Advance" to
DISP
continue with the calibration process.
JMP
1.008
ACMS
*
5790
*
HEAD
DCI References
PIC
552A410m
IEEE
OUT 3.2999mA, 0HZ; OPER; *OPC?[I][GTL]
IEEE
[D30000][GTL]
ACMS
5790
A
Figure 3-9. Sample MET/CAL Program
TOLERANCE
MOD1
U
C
switch is in NORMAL.
MOD2
3
4 CON
F
W
S
G
SH
N
2W