Dc Volts Calibration (Aux Output); Ac Volts Calibration (Aux Output) - Fluke 5520A Service Manual

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

5520A
Service Manual
# Retrieve AC-DC difference from data file named "A40-10mA"
1.059
1.060
1.061
1.061
1.062
1.063
# Correct the calculated value of AC Current by adding the AC-DC difference
# of the A40-series shunt used at the frequency under test
1.064
# Store corrected value into the UUT
1.065
1.066
1.067
1.068
# 'Ask' UUT for next value to calibrate
1.069

3-11. DC Volts Calibration (AUX Output)

To calibrate the auxiliary dc voltage function, use the same technique as previously
described for the normal dc voltage output, except use the AUX HI and LO terminals on
the UUT. Table 3-12 lists the calibration steps for AUX dc volts.
Step
1
2
3

3-12. AC Volts Calibration (AUX Output)

To calibrate the auxiliary ac voltage function, use the same technique as previously
described for the normal ac voltage output, except use the AUX HI and LO terminals on
the UUT. Table 3-13 lists the calibration steps for AUX dc volts.
Step
1
2
3
4
5
6
7
3-20
DOS
get_acdc A40-10mA
JMPT
1.064
OPBR
An error occurred during get_acdc
OPBR
Press YES to try again or NO to terminate.
JMPT
1.059
JMP
1.231
MATH
MEM = (M[21] * MEM) + M[21]
IEEE
CAL_NEXT [MEM]; *OPC?[I][GTL]
IEEE
ERR?[I$][GTL]
MATH
MEM1 = FLD(MEM2,1,",")
JMPT
1.231
IEEE
CAL_REF?[I][GTL]
Figure 3-9. Sample MET/CAL Program (cont)
Table 3-12. Calibration Steps for AUX DC Volts
Table 3-13. Calibration Steps for AUX Output AC Volts
Amplitude
300.000 mV
300.000 mV
3.00000 V
3.00000 V
5.0000 V
5.0000 V
3.0 V
5520A Output (AUX)
300.000 mV
3.00000 V
7.00000 V
5520A Output (AUX)
100 Hz
5 kHz
100 Hz
5 kHz
100 Hz
5 kHz
9.99 Hz
Frequency

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents