6-5.

Edge Specifications

Edge Characteristics into 50
Rise Time
Amplitude Range (p-p)
Resolution
Adjustment Range
Sequence Values
Frequency Range [1]
Typical Jitter, edge to trigger
Leading Edge Aberrations [2]
Typical Duty Cycle
Tunnel Diode Pulse Drive
[1] Above 2 MHz rise time specification < 350 ps.
[2] All edge aberration measurements made with Tektronix 11801 mainframe with SD26 input module.
Table 6-2. Edge Specifications
Load
300 ps
5.0 mV to 2.5 V
4 digits
10% around each sequence
value (indicated below)
5 mV, 10 mV, 25 mV, 50 mV, 60
mV, 80 mV, 100 mV, 200 mV,
250 mV, 300 mV, 500 mV, 600 mV,
1 V, 2.5 V
1 kHz to 10 MHz
< 5 ps (p-p)
within 2 ns from 50% of rising edge
2 to 5 ns
5 to 15 ns
after 15 ns
45% to 55%
Square wave at 100 Hz to 100 kHz, with variable amplitude of 60 V to
100 V p-p.
SC600 Option
SC600 Specifications
1-Year Absolute Uncertainty,
tcal
5 C
(+0 ps / -100 ps)
(2% of output + 200 V)
(2.5 ppm of setting)
< (3% of output + 2 mV)
< (2% of output + 2 mV)
< (1% of output + 2 mV)
< (0.5% of output + 2 mV)
6
6-7