Neuron-Spectrum-61, 62, 63, 64, 65 (Technical Manual)
Safety Measures
10
Do not use the system unless you read this manual and accompa-
nying documents.
The system is to be used by trained personnel only. No person
should attempt to use this system without necessary knowledge and
training to understand its use and how the results should be
interpreted. Before you start using the system read carefully this
manual, user manual for Neuron.Spectrum.NET software and
technical manuals for devices included into the system delivery set.
To ensure the safety and exclude the hazard of medical staff's
or patient's electric shock, the medical staff is PROHIBITED to:
 use the system which was mounted and installed improperly,
without following this manual instructions;
 connect the system and HF equipment to a patient simultane-
ously (it can cause burns in the places where the electrodes
are attached and it can also lead to the system damage);
 connect any devices not included in the system delivery set;
 open units included in the system delivery set;
 work with the system when the electronic unit, computer or
other devices used together are opened;
 connect electrodes placed on the patient to protective ground
or other conducting surfaces;
 apply the system in the highly explosive environment when
using SpO
channel (for example, if the flammable anesthetic
2
is present or inside the oxygen cabinet).
To ensure the safety of medical staff and patients it is PROHIBITED
to:
 immerse the system in water or any other solutions;
 use the system by children;
 drop or otherwise cause undue impact to the system. If the system
is dropped or otherwise damaged, return it to the manufacturer for
repair and/or check. Do not use the system if any damage is
suspected.