50 | DX-D 300 | Introduction
0172G EN 20230627 1651
WARNING:
The operator must take precautions to protect himself against
dangerous X-ray exposure when using the DR detector in the X-
ray beam path of an X-ray source.
CAUTION:
The DR Detector or CR cassette are not intended to be used as a
primary barrier to X-rays. The user is responsible for ensuring
the safety of the operator, bystanders, and the subjects being
radiographed.
CAUTION:
Strictly observe all warnings, cautions, notes and safety
markings within this document and on the product.
CAUTION:
All Agfa medical products must be used by trained and qualified
personnel.
CAUTION:
Avoid unnecessary dose by checking before exposure if the DR
Detector Switch displays the name of the DR Detector that is
being used and if the status of the DR Detector is ready for
exposure.
CAUTION:
System unavailability due to hardware or software failure. If the
product is used in critical clinical workflows, a backup system
has to be foreseen.
CAUTION:
Before moving the U-arm out of horizontal position, check that
no objects are lying on the bucky that can fall off.
CAUTION:
Damaged grid. Reduced image quality. Please handle the grids
with special care.
CAUTION:
When inserting the anti-scatter grids, it is essential that the grid
corresponds to the intended source-image-distance (SID) to
which the grid is focussed. Because of the focussing of the grids,
the tube head unit must be centered onto the bucky.
CAUTION:
Excessive ambient temperature may impact performance of DR
Detectors and cause permanent damage to the equipment. Refer
to the related user manual for environmental conditions for the
DR detector. If ambient temperature and humidity is outside the