Stopping the System

To stop the system:
1. Stop the MUSICA Acquisition workstation.
The MUSICA Acquisition workstation can be stopped in two ways, either
by logging out of Windows or without logging out of Windows.
For detailed information, refer to the MUSICA Acquisition workstation
User Manual, document 4420.
2. Switch off the U-arm unit using the button on the U-arm control box in the
examination room.
The green light goes out.
3. Press the Power OFF button on the X-ray generator control box to switch
off the generator.
The green light goes out.
4. In a configuration with a wireless DR detector, power off the DR detector:
• turn off the DR detector.
• remove the battery pack.
Note: Stopping the MUSICA Acquisition workstation does not
stop the DR detector. If the power of the DR Detector remains
on, no warming-up time will be needed after starting the
MUSICA Acquisition workstation.
Note: If the DR detector is powered down, a warming-up may be
required on the next start-up.
WARNING:
If the system has just been stopped, wait at least 10 seconds
before starting it again.
DX-D 300 | Operation | 71
0172G EN 20230627 1651