104 | DX-D 300 | Problem solving
Indicator
TECHNIQUE ERROR
PANEL OFF
(G)E63
0172G EN 20230627 1651
Description
Caution: If "TECHNIQUE ERROR" is
displayed during exposure it means
the exposure has been interrupted by
the "Security Timer" because of a
system failure. You should call field
service.
This error can also be shown if a failure on the auto-
matic collimator has been detected (blades are full
open or in movement during exposure, etc.). In this
case the indicator lights up continuously.
Indicates that the detector panel has been automati-
cally powered off due to:
• a power off in the X-ray generator control.
• an overheat (above 47 °C) in the detector panel.
Exposure with the DR detector is inhibited ex-
cept for the free exposure mode.
Excessive current in auxiliary winding during accel-
eration to 10000 RPM. Retry the exposure.