102 | DX-D 300 | Problem solving
Information message Description
Max mAs
Min ms
Max ms
Min ms & Min mA
Max ms & Max mA
Max Power
Space Charge
Max Energy (60kJ)
Instant Power
0172G EN 20230627 1651
Maximum mAs = 500 mAs (regulatory limit for
AEC)
Minimum exposure time (generator limit)
Maximum exposure time (generator limit)
Minimum exposure time and minimum mA (gener-
ator limit)
Maximum exposure time and maximum mA (gen-
erator limit)
Maximum power (generator limit)
Filament emission limit for a combination of kVp
and mA in the selected focal spot.
The maximum energy cannot exceed 60 kJ (kVp x
mAs = 60 kJ) (regulatory limit for AEC).
Instantaneous power limit of the X-ray tube de-
pending on the anode temperature, the selected ex-
posure time and the selected focal spot.