Magnetic field
30 A/m
at the supply
frequency
(50/60 Hz)
IEC 61000-4-8
• Note: U
is the alternating current in the network before the application
T
of the test level.
Table 17: Guidance and manufacturer's declaration on electromagnetic
immunity
Immunity Test
Radiated RF
3 Vrms
EM fields
from 80
IEC 61000-4-3
MHz to 2.7
GHz
(80% AM at
1 kHz)
Refer to the
Proximity
section "Im-
fields from RF
munity to
wireless com-
RF wireless
munications
communica-
equipment
tion equip-
IEC 61000-4-3
ment"
Conducted dis-
3 Vrms
turbances in-
from 150
duced by RF
kHz to 80
fields
Mhz
IEC 61000-4-6
6 Vrms in
ISM bands
from 150
kHz to 80
MHz
DX-D 300 | Remarks for HF-emission and immunity | 119
30 A/m
IEC
Compli-
60601-1-2
ance Level
Test Level
3 Vrms
from 80
MHz to 2.7
GHz
(80% AM
at 1 kHz)
Refer to
the section
"Immunity
to RF wire-
less com-
munica-
tion equip-
ment"
3 Vrms
from 150
kHz to 80
Mhz
6 Vrms in
ISM bands
from 150
kHz to 80
MHz
Magnetic field at the
network frequency
should correspond to
the typical values as
they are in a commer-
cial and clinical envi-
ronment.
Electromagnetic environment -
guidance
Portable RF communications
equipment (including peripher-
als such as antenna cables and
external antennas) should be
used no closer than 30 cm to
any part of the equipment, in-
cluding cables specified by
manufacturer. Otherwise, deg-
radation of the performance of
this equipment could result.