Dose Area Product Meter (Iba Dap) Technical Data - AGFA DX-D 300 User Manual

Hide thumbs Also See for DX-D 300:
Table of Contents

Advertisement

Dose Area Product Meter (IBA DAP) Technical Data

Manufacturer
Supported Type
Dose area product range
DAP resolution
Active area
Dimensions
Weight
Equivalent filtration of the ionization
chamber at 70 kV
Correction factors for using the DAP meter on high altitude
Environmental conditions
75 kPa (ca. 2500 m)
0° Celcius
75 kPa (ca. 2500 m)
20° Celcius
70 kPa (ca. 3000 m)
0° Celcius
70 kPa (ca. 3000 m)
20° Celcius
DX-D 300 | Technical Data | 115
IBA Dosimetry GmbH
Bahnhofstrasse 5
DE-90592 Schwarzenbruck
120-131 HS/RS485
(0.1...99999999.99) cGy x cm
2
0.01 cGy x cm
14.0 cm x 14.0 cm
17.9 cm x 16.6 cm x 1.7 cm (WxDxH)
approx. 220 g
0.31 mm Al
Correction factor
1.26
1.35
1.35
1.45
0172G EN 20230627 1651
2

Advertisement

Table of Contents
loading

Table of Contents