118 | DX-D 300 | Remarks for HF-emission and immunity
Table 16: Guidance and manufacturer's declaration on electromagnetic
immunity
Immunity
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short inter-
ruptions and
voltage varia-
tions on pow-
er supply in-
put lines.
IEC
61000-4-11
0172G EN 20230627 1651
IEC 60601-1-2
Test
Test Level
± 8 kV contact
± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV
air
± 2 kV for power
supply lines
± 1 kV for input/
output lines
(100 kHz repeti-
tion frequency)
± 0.5 kV, ± 1 kV
line(s) to line(s)
± 0.5 kV, ± 1 kV,
± 2 kV line(s) to
earth
0% U
for 0.5 cy-
T
cle at 0°, 45°,
90°, 135°, 180°,
225°, 270° and
315°
0% U
for 1 cycle
T
at 0°
70% U
T
25/30 cycles at
0% 250/300 cy-
cles
Compliance
Level
± 8 kV contact
± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV
air
± 2 kV for power
supply lines
± 1 kV for input/
output lines
(100 kHz repeti-
tion frequency)
± 0.5 kV, ± 1 kV
line(s) to line(s)
± 0.5 kV, ± 1 kV,
± 2 kV line(s) to
earth
0% U
for 0.5 cy-
T
cle at 0°, 45°,
90°, 135°, 180°,
225°, 270° and
315°
0% U
for 1 cycle
T
at 0°
for
70% U
T
25/30 cycles at
0% 250/300 cy-
cles
Electromagnetic en-
vironment - guid-
Floors should consist
of wood, concrete or
ceramic tiles. The rel-
ative humidity must
be at least 30%, if the
floor is made of syn-
thetic material.
The quality of the
voltage supplied
should correspond to
a typical commercial
or clinical environ-
ment.
The quality of the
voltage supplied
should correspond to
that of a typical com-
mercial or clinical en-
vironment.
The quality of the
voltage supply should
correspond to that of
a typical commercial
or clinical environ-
ment. If the user
wants the device to
work continuously,
for
even when the energy
supply is interrupted,
it is recommended to
use an energy supply
free of interruptions
or a battery.
ance